Visualization of pulse propagation with high frame rate spatial compound imaging

Ryo Nagaoka, Genta Masuno, Kazuto Kobayashi, Shin Yoshizawa, Shin ichiro Umemura, Yoshifumi Saijo

研究成果: Article査読

抄録

Shear wave Imaging is a technique that estimates tissue stiffness by measuring shear wave induced by Push Beam. However, a decay of estimate accuracy may accrue due to motions of such as pulsation and breathing. In this presentation, we propose estimation method of tissue stiffness based on the pulse propagation.Verasonic and 7.5 MHz Linear Probe were utilized to acquire raw data. The frame rate was 1 kfps. The identical data was used to measure the velocity by a modified 1D-cross-correlation-algorithm. A measurement area was right common carotid artery of 24-year-old healthy male.The pulse propagation was visualized with high-frame-rate ultrasound imaging. The dispersion-relation was observed. For example, the velocity of artery was 6.08 ± 0.19 m/s, and that of muscle was 7.43 ± 0.30 m/s at 15 Hz. It was possible to visualize the appearance of propagation induced by pulsation. The dispersion relation was revealed by the proposal method.

本文言語English
ページ(範囲)O-353-O-354
ジャーナルTransactions of Japanese Society for Medical and Biological Engineering
52
DOI
出版ステータスPublished - 2014 8 17

ASJC Scopus subject areas

  • 生体医工学