Usefulness of heidelberg Retina Tomograph II (version 3.0) in glaucoma screening

Yu Sawada, Makoto Ishikawa, Noriko Sato, Takeshi Yoshitomi

研究成果: Article査読

抄録

Purpose : To evaluate the usefulness of HRT II (Heidelberg Retina Tomograph II version 3.0) in public screening for glaucoma. Cases and Method : This study was made on 1,420 eyes of 710 persons who participated in glaucoma screening program. HRT II was persuaded and was processed by GPS global. Findings were compared with clinical diagnosis of glaucoma based on morphology of optic disc and visual field by Humphrey 24-2 test. Results : HRT was possible in 94.3% of participants, of which 82.9% gave reliable results. When compared with classification as glaucoma suspect by clinical diagnosis and guideline for borderline by HRT, the sensitivity ranged from 55.2 to 100% and specificity from 58.6 to 86.4%. Conclusion : HRT II processed by GPS global promises to be of value in detecting glaucoma in public screening program.

本文言語English
ページ(範囲)293-296
ページ数4
ジャーナルJapanese Journal of Clinical Ophthalmology
63
3
出版ステータスPublished - 2009
外部発表はい

ASJC Scopus subject areas

  • 眼科学

フィンガープリント

「Usefulness of heidelberg Retina Tomograph II (version 3.0) in glaucoma screening」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル