Tomographic quantitative phase measurement by hard X-ray micro-interferometer with 250 nm spatial resolution

Takahisa Koyama, Hidekazu Takano, Yoshiyuki Tsusaka, Yasushi Kagoshima

研究成果: Article

2 引用 (Scopus)

抜粋

Tomographic quantitative phase measurement has been carried out by using a hard X-ray micro-interferometer in Hyogo-ID at SPring-8. The micro-interferometer combines an imaging microscope and a wavefront-division-type interferometer by using a novel optical element called a "twin zone plate." A three-dimensional quantitative phase distribution of a glass capillary with an outer diameter of 6 μm was successfully obtained with the tomographic technique. The measured refractive index corresponded to the ideal value. The spatial resolution of the tomographic image was estimated to be 250 nm and the density resolution was estimated to be 0.6 g/cm3.

元の言語English
ページ(範囲)603-607
ページ数5
ジャーナルSpectrochimica Acta - Part B Atomic Spectroscopy
62
発行部数6-7 SPEC. ISS.
DOI
出版物ステータスPublished - 2007 7 1
外部発表Yes

ASJC Scopus subject areas

  • Analytical Chemistry
  • Atomic and Molecular Physics, and Optics
  • Instrumentation
  • Spectroscopy

フィンガープリント Tomographic quantitative phase measurement by hard X-ray micro-interferometer with 250 nm spatial resolution' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。