Temperature dependence of spectroscopic performance of thallium bromide X- and gamma-ray detectors

Toshiyuki Onoclera, Keitaro Hitomi, Taclayoshi Shoji

研究成果: Article査読

8 被引用数 (Scopus)

抄録

Thallium bromide (TlBr) is a compound semiconductor characterized by wide band-gap energy (2.68 eV), high atomic numbers (Tl: 81, Br: 35) and high density (7.56 g/cm3). Mobility-lifetime products for electrons ((μτ)e) and holes ((μτ)h) in TlBr crystals are 1 × 10-3 cm2/V and 3 × 10-4 cm2/V, respectively, which are comparable to μτ for carriers in CdTe crystals. Therefore, TlBr is a suitable material for fabrication of X- and gamma-ray detectors. In this study, TlBr detectors have been fabricated from crystals grown by the traveling molten zone method. The detectors have thicknesses of approximately 0.4 mm, and have gold electrodes (3 mm in diameter) on both surfaces of the crystals. In order to evaluate the temperature dependence of the charge transport properties of the detectors, (μτ) e and (μτ)h in the detectors have been estimated by the Hecht analysis at various temperatures of the detectors (-40,20,0,20 and 40°C). The (μτ)h in the detectors was of the order of 10-5 cm2/V at -40°C, and increased with increasing temperatures. The maximum value of the (μτ)h was obtained at 20°C, and was about ten times larger than the value at -40°C. On the other hand, the (μτ)e was approximately 1 × 10-3 cm 2/V between -40°C and 20°C. The temperature dependence of spectral response of the detectors has been evaluated by acquiring energy spectra of gamma-rays from 241Am at -40, 20, 0, 20 and 40°C. At -40°C, the full-energy peak corresponding to 59.5 keV gamma-rays has been obtained from the detectors, but the energy resolution of the peak was degraded due to the hole tailing on the low-energy side of the peak. The energy resolution of the detectors gradually improved with increasing temperatures of the detectors. The best FWHM energy resolution was measured to be 5.4 keV (9.1%) at 20°C.

本文言語English
ページ(範囲)860-863
ページ数4
ジャーナルIEEE Transactions on Nuclear Science
54
4
DOI
出版ステータスPublished - 2007 8

ASJC Scopus subject areas

  • 核物理学および高エネルギー物理学
  • 原子力エネルギーおよび原子力工学
  • 電子工学および電気工学

フィンガープリント

「Temperature dependence of spectroscopic performance of thallium bromide X- and gamma-ray detectors」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル