Small diameter electromagnetic single fiber endoscope using magnetic material tube and assist coil

Xuan Li, Tadao Matsnuaga, Yoichi Haga

研究成果: Article査読

抄録

The purpose of this study is to develop a small diameter (less than 1 mm) and high resolution endoscope. This probe consists of single optical fiber which is fixed a Ni tube and photo-fabricated metal saddle-shaped coils and solenoid coil which is fabricated by coiling machine. The optical fiber which is fixed a Ni tube is inserted into coil. When an alternating current supplies to the driving coils, the Ni tube is vibrated by magnetic force one dimensionally. With the assist of the magnetic force by additional solenoid shaped assist coils which are located in frontside and backside of the saddle-shaped coil, the vibration amplitude of the fiber is increased. The 1 mm diameter probe has been fabricated, and fiber vibration amplitude increase was confirmed.

本文言語English
ページ(範囲)O-495
ジャーナルTransactions of Japanese Society for Medical and Biological Engineering
52
DOI
出版ステータスPublished - 2014 8 17

ASJC Scopus subject areas

  • Biomedical Engineering

フィンガープリント 「Small diameter electromagnetic single fiber endoscope using magnetic material tube and assist coil」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル