Roundness and spindle error measurement by angular three-probe method (1st report)

Wei Gao, Eijiro Sato, Takamasa Ohnuma, Satoshi Kiyono

研究成果: Article査読

7 被引用数 (Scopus)

抄録

This paper presents a new error separation method for roundness and spindle error measurement, which is called the angular three-probe method. Three two-dimensional angle probes are employed in this method. The two-dimensional surface local slopes of the cylindrical workpiece, the radial error motion and angular error motion of the spindle, which are included in the output of each probe, can be separated from each other using the two-dimensional outputs of probes. Comparing with the conventional displacement three-probe method, the angular three-probe method is more suitable for detecting the multi-degrece-of-freedom components of spindle error and roundnesś. In this paper, the measurement of roundness is discussed in detail theoretically and experimentally. An in situ self-calibration method, which does not require extra angle references, is proposed to calibrate the angle probes. Experimental results have indicated that the roundness can be measured with a repeatability of approximately 0.04μm. Comparison of roundness measurement with the displacement three-probe method has also been conducted.

寄稿の翻訳タイトルRoundness and spindle error measurement by angular three-probe method (1st report)
本文言語Japanese
ページ(範囲)1195-1199
ページ数5
ジャーナルSeimitsu Kogaku Kaishi/Journal of the Japan Society for Precision Engineering
68
9
DOI
出版ステータスPublished - 2002 9
外部発表はい

Keywords

  • Angle probe
  • Angular three-probe method
  • Error separation
  • On-machine measurement
  • Roundness
  • Self-calibration

ASJC Scopus subject areas

  • Mechanical Engineering

フィンガープリント 「Roundness and spindle error measurement by angular three-probe method (1st report)」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。