Photoneutron yield in steel shield by electron beam from medical linear accelerator

Yukio Fujita, Hidetoshi Saitoh, Atsushi Myojoyama

    研究成果: Paper査読

    抄録

    In recent years, many medical linear accelerators do not equip a beam stopper and such linacs generate not only high energy photons but high energy electrons. Therefore, there is possibility of shield necessity against bremsstrahlen and photoneutron generated by high energy electron beam. However, there are a few physical data or no recommendation. In this report, yield efficiency and leakage dose of bremsstrahlen and photoneutron was calculated using EGS Monte Carlo simulations. Then, to verify result by calculation, leakage dose by photoneutron was measured with a rem counter. The result clearly shows that leakage dose of bremsstrahlen and photoneutron generated by 15 MeV electron beam is negligible.

    本文言語English
    ページ106-114
    ページ数9
    出版ステータスPublished - 2006 12 1
    イベント13th EGS Users' Meeting in Japan 2006 - Tsukuba, Japan
    継続期間: 2006 8 82006 8 10

    Other

    Other13th EGS Users' Meeting in Japan 2006
    CountryJapan
    CityTsukuba
    Period06/8/806/8/10

    ASJC Scopus subject areas

    • Software

    フィンガープリント 「Photoneutron yield in steel shield by electron beam from medical linear accelerator」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。