Performance of n-γ pulse-shape discrimination with simple pile-up rejection at high γ-ray count rates

T. Okuda, Hirohito Yamazaki, M. Kawabata, J. Kasagi, H. Harada

    研究成果: Article査読

    7 被引用数 (Scopus)

    抄録

    The performance of n-γ pulse-shape discrimination for a liquid scintillation detector has been investigated for γ-ray count rates up to 50 kcps. A method in which the ratio of the total to partial charge in the anode pulse is directly measured has shown much improved quality of the pulse-shape discrimination when pile-up events are rejected; it can discriminate neutron events of 50 cps from γ-ray events of 29 kcps. The method with simple pile-up rejection has the advantage that only general purpose electronics are required.

    本文言語English
    ページ(範囲)497-502
    ページ数6
    ジャーナルNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
    426
    2
    DOI
    出版ステータスPublished - 1999 5月 1