Optical phase contrast mapping of highly focused ultrasonic fields

Soichiro Harigane, Ryo Miyasaka, Shin Yoshizawa, Shin Ichiro Umemura

研究成果: Article査読

24 被引用数 (Scopus)

抄録

The most common method of measuring an ultrasonic pressure field is a hydrophone scan. However, this method has a long scanning time and disturbs the acoustic field. In this study, we used an optical phase contrast method for the measurement. Because this method uses light, fast and noninvasive measurement can be performed. The projections of an ultrasonic pressure field were obtained with a charge-coupled device (CCD) camera, and the three-dimensional (3D) acoustic pressure field was reconstructed using a computed tomography (CT) algorithm from these projections. The result was compared with that of hydrophone measurement and demonstrated the successful reconstruction of a focal ultrasonic pressure field.

本文言語English
論文番号07HF07
ジャーナルJapanese journal of applied physics
52
7 PART 2
DOI
出版ステータスPublished - 2013 7

ASJC Scopus subject areas

  • 工学(全般)
  • 物理学および天文学(全般)

フィンガープリント

「Optical phase contrast mapping of highly focused ultrasonic fields」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。