抄録
Compressed sensing magnetic resonance imaging (CS-MRI) uses random undersampling and nonlinear iterative reconstruction. This study was conducted to clarify the noise power spectrum (NPS) characteristics of CS-MRI. We measured two-dimensional (2D) NPS of CS-MRI with various acceleration factors (AF) and denoising factors (DF) and compared their appearance to those of conventional parallel MR images. Results showed that the 2D NPS of CS-MRI exhibited the following characteristics: (1) local decrease in the low-frequency region, (2) gradual decrease in the high-frequency region, and (3) a stripe pattern aligned at unequal intervals in the phase-encoding direction. Specifically, the 2D NPS of CS-MRI reflects the random undersampling pattern of k-space data. Additionally, 2D NPS allowed visualization of AF-dependent noise characteristics of CS-MRI. Furthermore, 1D NPS graph shapes clarified the CS-MRI noise characteristic dependence on AF and DF.
本文言語 | English |
---|---|
ページ(範囲) | 93-99 |
ページ数 | 7 |
ジャーナル | Radiological Physics and Technology |
巻 | 14 |
号 | 1 |
DOI | |
出版ステータス | Published - 2021 3月 |
外部発表 | はい |
ASJC Scopus subject areas
- 放射線
- 理学療法、スポーツ療法とリハビリテーション
- 放射線学、核医学およびイメージング