Microfocusing of 82 keV x rays with a sputtered-sliced Fresnel zone plate

M. Awaji, Y. Suzuki, A. Takeuchi, H. Takano, N. Kamijo, M. Yasumoto, Y. Terada, S. Tamura

研究成果: Article査読

6 被引用数 (Scopus)

抄録

An approach for microfocusing of 82 keV x-rays with a sputtered-sliced Fresnel zone plate (SS-FZP) was described. The SS-FZP consisted of 50 layers of alternative copper layer and aluminum layer deposited on a gold core. The measured distance between the SS-FZP and the focal point was 1.64 m at 82 keV. The achieved beam size was found to be 1.1 μm in vertical and 1.3 μm in horizontal directions while the measured flux density was 7×10 5 (photons/s/μm2).

本文言語English
ページ(範囲)4948-4949
ページ数2
ジャーナルReview of Scientific Instruments
74
11
DOI
出版ステータスPublished - 2003 11月
外部発表はい

ASJC Scopus subject areas