Local internal pressure measurement system of cerebral aneurysm model using ultra-miniature fiber-optic pressure sensor

Tadao Matsunaga, Akira Ito, Soyoka Osaki, Yasutomo Shimizu, Masaaki Shojima, Hiroshi Yoshida, Noriko Tsuruoka, Simon Tupin, Makoto Ohta, Yoichi Haga

研究成果: Conference contribution

1 被引用数 (Scopus)

抄録

Local internal pressure measurement system of cerebral aneurysm model has been developed for the quantitative evaluation of the therapeutic effects of the coil embolization. For local pressure measurement at multi point, an ultra-miniature fiber-optic pressure sensor, of which diameter is 125 μm and measurement system which has four channel optical switch has been developed. In this study, the cerebral aneurysm model equipped with two fiber-optic pressure sensors was fabricated and the pressure measurement of two sensors was performed.

本文言語English
ホスト出版物のタイトルMHS 2017 - 28th 2017 International Symposium on Micro-NanoMechatronics and Human Science
出版社Institute of Electrical and Electronics Engineers Inc.
ページ1-3
ページ数3
ISBN(電子版)9781538633144
DOI
出版ステータスPublished - 2018 2 28
イベント28th International Symposium on Micro-NanoMechatronics and Human Science, MHS 2017 - Nagoya, Japan
継続期間: 2017 12 32017 12 6

出版物シリーズ

名前MHS 2017 - 28th 2017 International Symposium on Micro-NanoMechatronics and Human Science
2018-January

Other

Other28th International Symposium on Micro-NanoMechatronics and Human Science, MHS 2017
国/地域Japan
CityNagoya
Period17/12/317/12/6

ASJC Scopus subject areas

  • バイオテクノロジー
  • 生体医工学
  • 電子工学および電気工学
  • 機械工学

フィンガープリント

「Local internal pressure measurement system of cerebral aneurysm model using ultra-miniature fiber-optic pressure sensor」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル