Initial organ distribution and biological safety of Mg 2+ released from a Mg alloy implant

Akimitsu Sato, Yoshinaka Shimizu, Yoshimichi Imai, Toshiji Mukai, Akiko Yamamoto, Chieko Miura, Kenji Muraki, Yuya Sano, Naoko Ikeo, Masahiro Tachi

研究成果: Article査読

3 被引用数 (Scopus)

抄録

Magnesium (Mg) alloys are considered promising materials for biodegradable medical devices; however, the initial effects and distribution of released Mg 2+ ions following implantation are unclear. This is addressed in the present study, using two types of Mg alloys implanted into rats. An in vitro immersion test was first carried out to quantify Mg 2+ ions released from the alloys at early stages. Based on these data, we performed an in vivo experiment in which large amounts of alloys were subcutaneously implanted into the backs of rats for 1, 5, 10, and 25 h. Mg 2+ accumulation in organs was measured by inductively coupled plasma mass spectrometry. In vivo, blood and urine Mg 2+ concentrations were higher in rats receiving the implants than in controls after 1 h; however, the levels were within clinically accepted guidelines. The Mg 2+ concentration in bone was significantly higher in the 25 h implanted group than in the other groups. Our results suggest that homeostasis is maintained by urinary excretion and bone accumulation of released Mg 2+ ions in response to sudden changes in Mg 2+ ion concentration in the body fluid in a large number of Mg alloy implants at the early stages.

本文言語English
論文番号035006
ジャーナルBiomedical Materials (Bristol)
13
3
DOI
出版ステータスPublished - 2018 3月 6

ASJC Scopus subject areas

  • バイオエンジニアリング
  • 生体材料
  • 生体医工学

フィンガープリント

「Initial organ distribution and biological safety of Mg 2+ released from a Mg alloy implant」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル