Improvement of transcutaneous communication system using intra-body communication for practical use

Eiji Okamoto, Kazuhide Miura, Yasuyuki Shiraishi, Tomoyuki Yambe, Yoshinori Mitamura

研究成果: Article

抜粋

We have been studying the transcutaneous communication system(TCS) using the intra-body communication, and we developed the implantable miniature TCS in which a Lithium battery was integrated on a circuit board. We examined the implantable TCS in the animal experiment, and the implanted TCS inside the goat’ thoracic cavity was not able to communication with the external unit at all. From the in vitro experiment for verification of the animal experiment showed that the amplitude of the receiving signal on the receiver is proposal to the distance between the battery and the circuit board on the transmitter, and integration of the battery on the transmitter circuit board is difficult to achieve data communication. From the results, we develop a new implantable miniature TCS unit in which the battery is separated from a circuit board. In addition, the receiver circuit is redesigned in order to improve receiving sensitivity.

元の言語English
ページ(範囲)42-43
ページ数2
ジャーナルTransactions of Japanese Society for Medical and Biological Engineering
56
発行部数Proc
DOI
出版物ステータスPublished - 2018 1 1

ASJC Scopus subject areas

  • Biomedical Engineering

フィンガープリント Improvement of transcutaneous communication system using intra-body communication for practical use' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。

  • これを引用