Applying downsized semiconductor detectors to PET cameras makes it possible to achieve high spatial resolutions. On the other hand, a difficulty in obtaining a high packing ratio of detectors arises from this downsizing. When detectors which construct a gantry are increased significantly according to downsizing, the volume of electronics becomes large and limits packing. Therefore, increase of detectors results in low packing ratio of detectors. As a solution of the packing problem, we propose using position sensitive type semiconductor detectors whose one of electrodes is a resistive plane. In such detectors, several charge collection electrodes are made on the resistive electrode and detection positions are identified by difference of collected charges among electrodes. Using such detectors contributes to decreasing electronics to be needed, and therefore an achievement of high packing can be expected. We report on attempts of two dimensional position sensitive CdTe detectors for a development of a PET camera using them. In our development, it is supposed that detector units of the camera are constructed by stacking two dimensional position sensitive semiconductor detectors whose thickness is small. Detection positions in the detector unit can be determined three-dimensionally by structure of the unit. A high spatial resolution in transaxial plane is guaranteed by thinness of the detectors. An axial detection position and a depth of interaction are specified by position sensitivity of the detector. Position resolution of the detector varies according to layout of charge collection electrodes. In this paper, some electrode layouts were attempted and evaluated with CdTe detectors (detector size: 20mm x20mm xl.Omm thickness).It was confirmed that a relation between actual detection position and calculated position was not linear. As an attempt to improve the un-linearity, we tried to connect external resistances among collection electrodes.

本文言語English
ホスト出版物のタイトル2008 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2008
ページ4924-4926
ページ数3
DOI
出版ステータスPublished - 2008
イベント2008 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2008 - Dresden, Germany
継続期間: 2008 10 192008 10 25

出版物シリーズ

名前IEEE Nuclear Science Symposium Conference Record
ISSN(印刷版)1095-7863

Other

Other2008 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2008
国/地域Germany
CityDresden
Period08/10/1908/10/25

ASJC Scopus subject areas

  • 放射線
  • 核物理学および高エネルギー物理学
  • 放射線学、核医学およびイメージング

フィンガープリント

「Fundamental study of two-dimensional position sensitive CdTe detector for PET camera」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル