Purpose: To present a novel method for analyzing the voltage waveform from high-frequency X-ray generators for radiographic systems. Material and Methods: The output signal of the actual voltage across the tube of a high-frequency generator was measured using the built-in voltage sense taps that are used for voltage regulation feedback in X-ray generators. The output signal was stored in an analyzing recorder, and the waveforms were analyzed using FFT analysis. The FFT analysis of high-frequency generators consisted of obtaining the power spectrum, comparing the major frequency components in the tube voltage waveforms, and examining the intensity of each frequency component. Results: FFT analysis enables an objective comparison of the complex tube voltage waveforms in high-frequency X-ray generators. FFT analysis detected the change in the X-ray tube voltage waveform that occurred when there were problems with the high-frequency generator. Conclusion: High-frequency X-ray generators are becoming the universal choice for radiographic systems. The X-ray tube voltage and its waveform are important features of an X-ray generator, and quality assurance (QA) is important, too. As a tool for engineers involved in the design and development of X-ray generators, we can see that our methods (FFT analysis) might have some value as a means of describing generator performance under varying conditions. Furthermore, since the X-ray tube voltage waveform of a high-frequency generator is complex, FFT analysis may be useful for QA of the waveform.

元の言語English
ページ(範囲)810-814
ページ数5
ジャーナルActa Radiologica
46
発行部数8
DOI
出版物ステータスPublished - 2005 12 1
外部発表Yes

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

フィンガープリント FFT analysis of the X-ray tube voltage waveforms of high-frequency generators for radiographic systems' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。

  • これを引用