Development of a control system for pulsed-electron spin resonance spectrometers

Yukio Mizuta, Shunji Kazama, Yasunori Ohba, Naoko Sakai, Yukio Yamamoto, Yuhei Shimoyama

研究成果: Article

5 引用 (Scopus)

抜粋

A pulse control and data acquisition system with high time resolution of 300 ps and data-averaging rate of 100 kHz for a pulsed-ESR spectrometer was constructed. The system consists of two sets of 3.34 GHz data timing generator and a dual-channel analog-to-digital converter with 1 GHz samples, both of which are commercially available. We developed a scheme and software to control these equipments with high efficiency in pulsing design, data acquisition, and accurate synchronization of the devices. The system was developed at Ku band (17.5 GHz) and demonstrated some examples of electron spin echo and double quantum measurements.

元の言語English
記事番号044705
ジャーナルReview of Scientific Instruments
79
発行部数4
DOI
出版物ステータスPublished - 2008 5 8