Determination of the detection efficiency of a channelplate electron multiplier

Toshio Sakurai, T. Hashizume

    研究成果: Article査読

    25 被引用数 (Scopus)

    抄録

    Making use of a technique of layer-by-layer field evaporation developed in the time-of-flight atom probe, the detection efficiency of a microchannelplate supplied by Galileo Electro Optics is determined. The detection efficiency for single ion impact is 57 ± 8%, close to the fraction of the active channel area of the surface. It increases slightly, but no more than a few percent, even the front surface is positively biased.

    本文言語English
    ページ(範囲)236-239
    ページ数4
    ジャーナルReview of Scientific Instruments
    57
    2
    DOI
    出版ステータスPublished - 1986 12 1