抄録
In order to evaluate the analytical precision in low pressure laser-induced argon plasma emission spectrometry (LIPS), we conducted the determination of Ni in Fe-based alloys to consider the most suitable measurement condition, in particular, a gas pressure of argon and several measuring parameters for time-resolved observation. The selection of emission lines was an important factor to obtain a liner calibration curve depending on the concentration of Ni. The precision was generally inferior to analytical results in XRF and ICP-OES. However, the analytical value of Ni was obtained with good reliability, by averaging during replicate measurements. It was further shown the Ni content in Fe-Ni-X alloys would be determined by using calibration curves estimated from binary Fe-Ni alloys.
本文言語 | English |
---|---|
ページ(範囲) | 63-68 |
ページ数 | 6 |
ジャーナル | Tetsu-To-Hagane/Journal of the Iron and Steel Institute of Japan |
巻 | 98 |
号 | 2 |
DOI | |
出版ステータス | Published - 2012 |
ASJC Scopus subject areas
- 凝縮系物理学
- 物理化学および理論化学
- 金属および合金
- 材料化学