A New Microstrip Pickup Coil for Thin-Film Permeance Meters

T. Kawazu, M. Yamaguchi, K. I. Arai, T. Kawazu, M. Yamaguchi, K. I. Arai, M. Yamaguchi

研究成果: Article査読

22 被引用数 (Scopus)

抄録

In this paper we propose an application of microstrip line for a single-turn pickup coil in the thin-film permeance meters. This coil has two electric ports with one terminated into 50 Ω. It is free from LC resonance up to a few GHz because the microstrip line does not resonates and because the stray capacitance between the pickup coil and the driving sheets is small as compared with the conventional figure-8 coil. We measured the thin film permeance up to 700 MHz which is the upper frequency limit of the driving sheets used. The proposed coil has a potential to measure thin-film permeances over the wide range from 1 MHz up to a few GHz.

本文言語English
ページ(範囲)4641-4643
ページ数3
ジャーナルIEEE Transactions on Magnetics
30
6
DOI
出版ステータスPublished - 1994 11

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Electrical and Electronic Engineering

フィンガープリント 「A New Microstrip Pickup Coil for Thin-Film Permeance Meters」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。