TH-7B pneumatically driven sac-type ventricular assist device and drive system.

S. Nitta, T. Yambe, Y. Katahira, T. Sonobe, S. Naganuma, H. Akiho, S. Chiba, Y. Kakinuma, H. Matsuzawa, H. Hayashi

Research output: Contribution to journalArticlepeer-review

Abstract

TH-7B pneumatically driven sac-type ventricular assist device and drive system including automatic control system were developed and evaluated in the chronic animal experiments. The blood-contacting surfaces of the blood pump are coated with either Cardiothane or Cardiomat. In the chronic experiments using adult goats, only one of fifteen goats showed sudden thrombus formation after 34 pumping days. No gross findings of thrombus formation were observed in any of the cases, except for a thin circular thrombus at the junction of the connectors and the sac of the blood pump. These results suggest that antithrombogenicity of this system was satisfactory for clinical use.

Original languageEnglish
Pages (from-to)19-27
Number of pages9
JournalThe science reports of the research institutes, Tohoku University. Ser. C, Medicine. Tōhoku Daigaku
Volume38
Issue number1
Publication statusPublished - 1991 Jul

ASJC Scopus subject areas

  • Medicine(all)

Fingerprint Dive into the research topics of 'TH-7B pneumatically driven sac-type ventricular assist device and drive system.'. Together they form a unique fingerprint.