Abstract

Elucidating causes of infertility is important to conduct proper treatment for infertility. Mechanical characteristics of the endometrium may be mentioned as a cause of infertility. In this paper, we focus on softness of the endometrium. As a fundamental research for the softness measurement of the endometrium, we developed a system for measuring softness of cylindrical objects by using a balloon catheter. The measurement system is composed of a balloon catheter, a syringe and a pressure sensor. Young's modulus of cylindrical objects is calculated from balloon radius and internal pressure when the balloon is expanded in the object. In the experiment, we prepared four kinds of cylindrical samples with different softness. By using the measurement system, Young's modulus of each cylindrical sample is estimated. And the estimated values are compared with Young's modulus measured by compression test. From the results, it was found that the developed system is effective for measuring softness of cylindrical objects.

Original languageEnglish
Pages (from-to)1127-1134
Number of pages8
JournalInternational Journal of Applied Electromagnetics and Mechanics
Volume59
Issue number3
DOIs
Publication statusPublished - 2019 Jan 1

Keywords

  • Balloon catheter
  • measurement
  • softness
  • young's modulus

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Condensed Matter Physics
  • Mechanics of Materials
  • Mechanical Engineering
  • Electrical and Electronic Engineering

Fingerprint Dive into the research topics of 'Softness measurement system for biological object by using balloon catheter'. Together they form a unique fingerprint.

  • Cite this