T1 - [Reduction method of patients' radiation dose considering the size of field of view with a digital cine X-ray system loading a flat-panel detector].
AU - Kato, Mamoru
AU - Chida, Koichi
AU - Yoshida, Kyouhei
AU - Sasaki, Fumiaki
AU - Sasaki, Masafumi
AU - Oosaka, Hajime
AU - Tosa, Tetsuo
PY - 2011
Y1 - 2011
N2 - Angiographic equipment has tended toward using flat-panel detector (FPD). However, the radiation dose varies depending on the size of field of view (FOV) in the FPD. Although we were using 7 inch FOV in cardiac procedures, we examined a way to reduce the patients' radiation dose by using 8 inch FOV utilizing a digital zoom processing. We measured the absorbed dose rates at an interventional reference point (IRP) with 20 cm acrylic board to compare 8 inch FOV using a digital zoom processing with 7 inch FOV. Additionally we visually evaluated these phantom images. Result of comparison of absorbed dose rates at IRP, 8 inch FOV using a digital zoom processing was reduced more than 7 inch FOV (fluoroscopic 29.0% and digital cine 37.5%). These phantom images were of equivalent quality. The method to use 8 inch FOV utilizing a digital zoom processing technique in cardiac procedures is one of the effective radiation exposure reduction methods.
AB - Angiographic equipment has tended toward using flat-panel detector (FPD). However, the radiation dose varies depending on the size of field of view (FOV) in the FPD. Although we were using 7 inch FOV in cardiac procedures, we examined a way to reduce the patients' radiation dose by using 8 inch FOV utilizing a digital zoom processing. We measured the absorbed dose rates at an interventional reference point (IRP) with 20 cm acrylic board to compare 8 inch FOV using a digital zoom processing with 7 inch FOV. Additionally we visually evaluated these phantom images. Result of comparison of absorbed dose rates at IRP, 8 inch FOV using a digital zoom processing was reduced more than 7 inch FOV (fluoroscopic 29.0% and digital cine 37.5%). These phantom images were of equivalent quality. The method to use 8 inch FOV utilizing a digital zoom processing technique in cardiac procedures is one of the effective radiation exposure reduction methods.
UR - http://www.scopus.com/inward/record.url?scp=84857941142&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84857941142&partnerID=8YFLogxK
U2 - 10.6009/jjrt.67.1443
DO - 10.6009/jjrt.67.1443
M3 - Article
C2 - 22104237
AN - SCOPUS:84857941142
SN - 0369-4305
VL - 67
SP - 1443
EP - 1447
JO - Nihon Hoshasen Gijutsu Gakkai zasshi
JF - Nihon Hoshasen Gijutsu Gakkai zasshi
IS - 11
ER -