Reducing the weight of mounted system for direct feeding FES

Kanzumasa Kan, Tadakuni Sato, Tetsuya Takura, Fumihiro Sato, Hidetoshi Matsuki

Research output: Contribution to journalArticle

Abstract

FES is the therapy for rehabilitate lost movement function to apply electrical stimulation to paralyzed extremity. We adapted implanted direct feeding method for FES. Direct feeding method is the stimulus method that transmits the energy for driving implants and communicates bi-directionally by electromagnetic induction with a mounted coil and implant coils contactlessly.However, we guess mounted system is too heavy for paralyzed patients. So we must reduce the weight.We proposed the new mounted coil with plural repeaters, the circuit composed of a capacitor and a coil with series connection. We designed mounted coil that can drive implanted coil by controlling a current in each repeater and magnetic field generated from a mounted coil.Consequently, we realized the 18turns lighter mounted exciting coil than the traditional coil and confirmed this coil can generate necessary magnetic field, 0.35mT, to implanted coils.

Original languageEnglish
Pages (from-to)O-515-O-516
JournalTransactions of Japanese Society for Medical and Biological Engineering
Volume52
DOIs
Publication statusPublished - 2014 Aug 17

Keywords

  • Contactless energy supply
  • Functional electrical stimulation
  • Implant
  • Paralysis
  • Repeater

ASJC Scopus subject areas

  • Biomedical Engineering

Fingerprint Dive into the research topics of 'Reducing the weight of mounted system for direct feeding FES'. Together they form a unique fingerprint.

  • Cite this