Preliminary study of using imaging plates to map skin dose of patients in interventional radiology procedures

Hiroko Ohuchi, T. Satoh, Y. Eguchi, K. Mori

Research output: Contribution to journalArticlepeer-review

5 Citations (Scopus)

Abstract

A method using europium-doped BaFBr imaging plates (IPs) has been studied for mapping entrance skin doses during interventional radiology (IR); the mapping is useful for detecting overlap between irradiation fields and determining the most exposed skin areas. IPs, which are two-dimensional radiation sensors made of photostimulated luminescence materials, have a linear dose response up to ∼100 Gy, can accurately measure doses from 1 μGy to 10 Gy and can be used repeatedly. Because the energy dependence of IPs is rather high, the IPs were characterised in this study and a sensitivity variation of ∼13% was observed for effective energies of 32.7 to 44.7 keV, which are used in IR procedures. Simulation of actual interventional cardiology procedures showed that the variation of sensitivity was within 5%, meaning that IPs are practical for measuring skin doses during IR. Moreover, the patient data can be stored online and easily called up when IR procedures must be repeated, helping to prevent radiation injuries.

Original languageEnglish
Pages (from-to)432-439
Number of pages8
JournalRadiation protection dosimetry
Volume117
Issue number4
DOIs
Publication statusPublished - 2005 Dec

ASJC Scopus subject areas

  • Radiation
  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging
  • Public Health, Environmental and Occupational Health

Fingerprint Dive into the research topics of 'Preliminary study of using imaging plates to map skin dose of patients in interventional radiology procedures'. Together they form a unique fingerprint.

Cite this