Performance of the LaBr3(Ce) scintillator for nuclear resonance fluorescence experiment

Toshiteru Kii, Kai Masuda, Hideaki Ohgaki, Hideo Harada, Fumito Kitatani, Takehito Hayakawa, Toshiyuki Shizuma, Nobuhiro Kikuzawa, Ryoichi Hajima, Nobuyuki Nishimori, Hiroyuki Toyokawa

Research output: Chapter in Book/Report/Conference proceedingConference contribution

7 Citations (Scopus)

Abstract

Performance of the LaBr3(Ce) scintillator for nuclear resonance fluorescence (NRF) experiment was investigated. Linearity and energy resolutions of the LaBr3(Ce) scintillator of 3 inch in diameter x 3 inch in thickness were measured using NRF scattering gamma ray from 208Pb (5.5 MeV and 7.3 MeV) and 24Mg (10.7 MeV). The linearity of the output for the gamma ray energy was less than 1 % and the energy resolution of the scintillator in keV was almost proportional to the square root of the gamma ray energy up to 10.7 MeV.

Original languageEnglish
Title of host publication2009 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2009
Pages1490-1492
Number of pages3
DOIs
Publication statusPublished - 2009 Dec 1
Event2009 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2009 - Orlando, FL, United States
Duration: 2009 Oct 252009 Oct 31

Publication series

NameIEEE Nuclear Science Symposium Conference Record
ISSN (Print)1095-7863

Other

Other2009 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2009
CountryUnited States
CityOrlando, FL
Period09/10/2509/10/31

ASJC Scopus subject areas