Monochromator for a soft X-ray photochemistry beamline BL27SU of SPring-8

H. Ohashi, E. Ishiguro, Y. Tamenori, H. Okumura, A. Hiraya, H. Yoshida, Y. Senba, K. Okada, N. Saito, I. H. Suzuki, K. Ueda, T. Ibuki, S. Nagaoka, I. Koyano, T. Ishikawa

Research output: Contribution to journalArticle

208 Citations (Scopus)

Abstract

A high-resolution monochromator with varied line space plane gratings (VLSG) and spherical focusing mirrors was installed in one of three branches of BL27SU in SPring-8. The performance of the monochromator was roughly evaluated from the photo ion yield of nitrogen molecule. Furthermore, the kinetic energy of the photoelectron from Xe 5p3/2 orbit was also measured at the same photon energy with the N2 to avoid the influence of natural width. The resolving power over 104 has been confirmed at the N K-edge.

Original languageEnglish
Pages (from-to)533-536
Number of pages4
JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume467-468
DOIs
Publication statusPublished - 2001 Jul 21

Keywords

  • Beamline
  • Flux
  • Monochromator
  • Nitrogen
  • Photoabsorption spectrum
  • Photochemistry
  • Resolving power
  • SPring-8
  • Soft X-ray
  • Synchrotron radiation

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Instrumentation

Fingerprint Dive into the research topics of 'Monochromator for a soft X-ray photochemistry beamline BL27SU of SPring-8'. Together they form a unique fingerprint.

  • Cite this

    Ohashi, H., Ishiguro, E., Tamenori, Y., Okumura, H., Hiraya, A., Yoshida, H., Senba, Y., Okada, K., Saito, N., Suzuki, I. H., Ueda, K., Ibuki, T., Nagaoka, S., Koyano, I., & Ishikawa, T. (2001). Monochromator for a soft X-ray photochemistry beamline BL27SU of SPring-8. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 467-468, 533-536. https://doi.org/10.1016/S0168-9002(01)00405-3