    Abstract

    The plasma of several different gases has shown a sporicidal activity. From these gases, nitrogen gas was most difficult to produce atomic nitrogen radicals. However, these radicals have a high energy, indicating that nitrogen gas plasma could be used to sterilize microorganisms and inactivate endotoxins. The sterilization mechanism of nitrogen gas plasma is the synergistic effect of a high rising-up voltage pulse, UV irradiation and atomic nitrogen radicals. Thus, the target cells were damaged by degradation, which resulted in death. The biological indicator (BI) used in this study was Geobacillus stearothermophilus ATCC 7953 at a population of 1 × 106 CFU/sheet. Sterility assurance was confirmed by using the BI. Moreover, endotoxins were successfully inactivated. More than 5 log reduction of endotoxins could be attained with 30 minutes of nitrogen gas plasma exposure. Material functionality influenced by nitrogen gas plasma presented a satisfactory result. No deterioration of polymers could be observed by nitrogen gas plasma exposure.

    Original languageEnglish
    Pages (from-to)131-143
    Number of pages13
    JournalBiocontrol Science
    Volume12
    Issue number4
    DOIs
    Publication statusPublished - 2007 Dec

    Keywords

    • Endotoxin
    • Inactivation
    • Microorganism
    • Nitrogen gas plasma
    • Sterilization

    ASJC Scopus subject areas

    • Applied Microbiology and Biotechnology
    • Public Health, Environmental and Occupational Health

    Fingerprint Dive into the research topics of 'Inactivation of microorganisms and endotoxins by low temperature nitrogen gas plasma exposure'. Together they form a unique fingerprint.

    Cite this