In vitro performance of trans-valve left ventricular assist device installed at aortic valve position

Eiji Okamoto, Tetsuya Yano, Yusuke Inoue, Yasuyuki Shiraishi, Tomoyuki Yambe, Yoshinori Mitamura

Research output: Contribution to journalArticle

Abstract

In this study, we developed a trans-valve left ventricular assist device (LVAD) that unites a rear-impeller axial-flow blood pump (AFBP) and a polymer membrane valve placed at the aortic valve position. The diameter and length of the rear impeller AFBP was 12 and 63 mm, respectively. The polymer membrane valve was similar to the jelly-fish valve consisting of a valve leaflet made of silicone rubber (thickness 0.5 mm), valve ring (diameter: 25 mm), and valve spokes. The trans-valve LVAD was examined in a mock circulation. An implantable pulsatile flow (PF) VAD was connected to an atrial reservoir to simulate the left ventricle (LV), and the Hall valve was worn in the inflow port, and the trans-valve LVAD was placed in the outflow port as an outflow valve. When the motor rotational speed increased to 26 400 rpm, the mean aortic flow increased from 4.2 to 5.3 L/min, mean aortic pressure increased from 83.4 to 100 mm Hg, and mean motor current of the implantable PF VAD decreased from 1.18 to 0.94 A (unloading effect on LV −21%). The energy equivalent pressure increased from 85.2 to 102 mm Hg, and surplus hemodynamic energy (SHE) decreased by −15.4% from the baseline. In conclusion, the trans-valve LVAD has an advantage of preserving pulsatility without any complicated mechanism and is a novel and promising LV support device.

Original languageEnglish
Pages (from-to)1067-1072
Number of pages6
JournalArtificial Organs
Volume44
Issue number10
DOIs
Publication statusAccepted/In press - 2020

Keywords

  • aortic valve
  • artificial heart
  • axial-flow blood pump
  • left ventricular assist device
  • valvo-pump

ASJC Scopus subject areas

  • Bioengineering
  • Medicine (miscellaneous)
  • Biomaterials
  • Biomedical Engineering