Abstract
We propose a new method for improving the characteristics of the piezoelectric transducer using by the inverse matrix of the cascade system so that time resolution in the determination of transit time of ultrasound signals in a thin layer is increased. The complete set of characteristic of the pulse echo system is described by a cascade of distributed constant systems for the ultrasonic transducer, matching layer, and acoustic medium. Acoustic input impedance and reflection characteristics are estimated by using the inverse matrix of the cascade system and the voltage and the current signal at the electrical port. From the inverse Fourier transform of the resultant reflection characteristic, transit time in an object that forms a thin layer is accurately determined with high time resolution. The principle of the method was confirmed by simulation experiments and experiments using a water tank.
Original language | English |
---|---|
Pages (from-to) | 3-14 |
Number of pages | 12 |
Journal | Japanese Journal of Medical Ultrasonics |
Volume | 23 |
Issue number | 6 |
Publication status | Published - 1996 Jan 1 |
Keywords
- 2- Terminal-pair network
- Acoustic input impedance
- Cascade matrix
- Reflection characteristic
- Ultrasonic transducer