N2 - We developed a new method of grazing exit electron probe microanalysis (GE-EPMA) and applied it to analyze both Si surfaces and Mg-salt particles. In conventional EPMA, X-rays are detected at an exit (takeoff) angle of approximately 45°. Therefore, when particles collected on a sample carrier are analyzed by EPMA, the X-rays from both the particles and the carrier are detected, although we need only the X-rays emitted form the particle itself. In contrast to this, the X-rays are detected at grazing exit angles in GE- EPMA. The X-rays emitted from deep inside of the sample are not detected under grazing exit conditions, and only X-rays emitted from the surface and the particle are measured. It was found that surface-sensitive analysis of a Si wafer was possible with low background at grazing exit angles. The intensity ratio of O Kα to Si Kα increased near zero degrees, indicating that the Si wafer is covered with a native Si oxide. Moreover, Mg Kα X-rays from a Mg-salt particle, which was deposited on the Si wafer, were detected with a small Si Kα intensity at grazing exit angles of less than 0.5°. By decreasing the exit angle to less than zero, only the top of the particle was observed; therefore, GE-EPMA measurement would make it possible to investigate the surface layer of one particle.
AB - We developed a new method of grazing exit electron probe microanalysis (GE-EPMA) and applied it to analyze both Si surfaces and Mg-salt particles. In conventional EPMA, X-rays are detected at an exit (takeoff) angle of approximately 45°. Therefore, when particles collected on a sample carrier are analyzed by EPMA, the X-rays from both the particles and the carrier are detected, although we need only the X-rays emitted form the particle itself. In contrast to this, the X-rays are detected at grazing exit angles in GE- EPMA. The X-rays emitted from deep inside of the sample are not detected under grazing exit conditions, and only X-rays emitted from the surface and the particle are measured. It was found that surface-sensitive analysis of a Si wafer was possible with low background at grazing exit angles. The intensity ratio of O Kα to Si Kα increased near zero degrees, indicating that the Si wafer is covered with a native Si oxide. Moreover, Mg Kα X-rays from a Mg-salt particle, which was deposited on the Si wafer, were detected with a small Si Kα intensity at grazing exit angles of less than 0.5°. By decreasing the exit angle to less than zero, only the top of the particle was observed; therefore, GE-EPMA measurement would make it possible to investigate the surface layer of one particle.
UR - http://www.scopus.com/inward/record.url?scp=0033168126&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0033168126&partnerID=8YFLogxK
U2 - 10.1021/ac990075p
DO - 10.1021/ac990075p
M3 - Article
C2 - 21662793
AN - SCOPUS:0033168126
VL - 71
SP - 2497
EP - 2501
JO - Industrial And Engineering Chemistry Analytical Edition
JF - Industrial And Engineering Chemistry Analytical Edition
SN - 0003-2700
IS - 13
ER -