T1 - Evaluation of implant screw loosening by resonance frequency analysis with triaxial piezoelectric pick-up
T2 - in vitro model and in vivo animal study
AU - Miyashita, Makiko
AU - Ogawa, Toru
AU - Naito, Hideki
AU - Shibamoto, Aya
AU - Wang, Angelina Shuhua
AU - Shobara, Kenta
AU - Sasaki, Keiichi
N1 - Funding Information:
Funding This study was supported by a Grant-in-Aid for Challenging Exploratory Research from the Ministry of Education, Culture, Sports, Science and Technology, Japan (grant no. 26670830).
Publisher Copyright:
© 2018, Springer-Verlag GmbH Germany, part of Springer Nature.
PY - 2018/6/1
Y1 - 2018/6/1
N2 - Objective: The aim of this study was to evaluate implant screw loosening using resonance frequency (RF) analysis with triaxial piezoelectric pick-up in vitro and in vivo. Methods: For the in vitro experiment, a titanium implant was inserted into a mandibular model. The abutment screws were tightened to 10 N torque and loosened from 0 to 90° for RF measurement using 13 different screw conditions. In the in vivo experiment, three titanium implants were installed in each tibia of a New Zealand white rabbit, and the RF values were recorded after 8 weeks. For the RF analyses, a small 3D accelerometer was mounted rigidly onto the implant abutment, and impulsive vibration was directly applied to the abutment to vibrate the implant in a direction perpendicular to the tibia and implant (x-axis). The y-axis was defined as parallel to the tibia. The RF values of the x- and y-axis directions (RF-X and RF-Y) were used for evaluation. Results: The RF values significantly decreased according to the degree of screw loosening (p < 0.05, ANOVA). In the in vitro model, RF-Y with x- and y-axis vibrations (RF-Yx and RF-Yy) significantly differed from the initial value at 10 and 15°, respectively (p < 0.05, Dunnett’s test). In the in vivo experiment, the RF-Yy significantly differed from the initial value between 5 and 20° (p < 0.05). Conclusion: The results suggest that RF analysis with triaxial piezoelectric pick-up can be used to detect implant screw loosening. Clinical relevance: RF analysis with the triaxial piezoelectric pick-up can be used to detect screw loosening after mounting the superstructure.
AB - Objective: The aim of this study was to evaluate implant screw loosening using resonance frequency (RF) analysis with triaxial piezoelectric pick-up in vitro and in vivo. Methods: For the in vitro experiment, a titanium implant was inserted into a mandibular model. The abutment screws were tightened to 10 N torque and loosened from 0 to 90° for RF measurement using 13 different screw conditions. In the in vivo experiment, three titanium implants were installed in each tibia of a New Zealand white rabbit, and the RF values were recorded after 8 weeks. For the RF analyses, a small 3D accelerometer was mounted rigidly onto the implant abutment, and impulsive vibration was directly applied to the abutment to vibrate the implant in a direction perpendicular to the tibia and implant (x-axis). The y-axis was defined as parallel to the tibia. The RF values of the x- and y-axis directions (RF-X and RF-Y) were used for evaluation. Results: The RF values significantly decreased according to the degree of screw loosening (p < 0.05, ANOVA). In the in vitro model, RF-Y with x- and y-axis vibrations (RF-Yx and RF-Yy) significantly differed from the initial value at 10 and 15°, respectively (p < 0.05, Dunnett’s test). In the in vivo experiment, the RF-Yy significantly differed from the initial value between 5 and 20° (p < 0.05). Conclusion: The results suggest that RF analysis with triaxial piezoelectric pick-up can be used to detect implant screw loosening. Clinical relevance: RF analysis with the triaxial piezoelectric pick-up can be used to detect screw loosening after mounting the superstructure.
KW - Animal study
KW - Dental implant
KW - Resonance frequency analysis
KW - Screw loosening
UR - http://www.scopus.com/inward/record.url?scp=85046040332&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85046040332&partnerID=8YFLogxK
U2 - 10.1007/s00784-018-2467-1
DO - 10.1007/s00784-018-2467-1
M3 - Article
C2 - 29705965
AN - SCOPUS:85046040332
SN - 1432-6981
VL - 22
SP - 2129
EP - 2134
JO - Clinical Oral Investigations
JF - Clinical Oral Investigations
IS - 5
ER -