TY - JOUR
T1 - Evaluation of a novel automated chemiluminescent assay system for antimicrobial susceptibility testing
AU - Manome, Isao
AU - Ikedo, Masanari
AU - Saito, Yoshihiko
AU - Ishii, Keiko Kumura
AU - Kaku, Mitsuo
PY - 2003/1/1
Y1 - 2003/1/1
N2 - The newly developed Rapid Lumi Eiken/IS60 (RL/IS60) system automatically determines MICs by detecting chemiluminescence produced in the reaction of a chemiluminescent probe and oxygen metabolites from living microorganisms. The present study evaluated this system for accuracy in antimicrobial susceptibility testing. Chemiluminescence intensities after 4 h of cultivation of clinically important strains were plotted against various concentrations of antimicrobial agents, which resulted in curves reflecting the levels of susceptibility. Sixty-percent inhibitory concentrations based on the susceptibility curves agreed with MICs determined by the reference microdilution method. When the MICs of antimicrobial agents for four quality control (QC) strains (Staphylococcus aureus, Enterococcus faecalis, Escherichia coli, and Pseudomonas aeruginosa) were determined by the RL/IS60 system, most (91.1%) of them were within the QC limits proposed by the National Committee for Clinical Laboratory Standards. The system was further assessed for a total of 162 clinical isolates, including E. coli, Citrobacter freundii, Enterobacter cloacae, Klebsiella pneumoniae, Serratia marcescens, Proteus mirabilis, Morganella morganii, P. aeruginosa, Haemophilus influenzae, S. aureus, coagulase-negative staphylococci, Enterococcus faecalis, Enterococcus faecium, and Streptococcus pneumoniae. Overall, there was 90.6% agreement between the RL/IS60 system and the reference microdilution method. Our results suggest that the RL/IS60 system provides rapid and reliable MICs of a variety of antimicrobial agents for clinical isolates as well as QC strains.
AB - The newly developed Rapid Lumi Eiken/IS60 (RL/IS60) system automatically determines MICs by detecting chemiluminescence produced in the reaction of a chemiluminescent probe and oxygen metabolites from living microorganisms. The present study evaluated this system for accuracy in antimicrobial susceptibility testing. Chemiluminescence intensities after 4 h of cultivation of clinically important strains were plotted against various concentrations of antimicrobial agents, which resulted in curves reflecting the levels of susceptibility. Sixty-percent inhibitory concentrations based on the susceptibility curves agreed with MICs determined by the reference microdilution method. When the MICs of antimicrobial agents for four quality control (QC) strains (Staphylococcus aureus, Enterococcus faecalis, Escherichia coli, and Pseudomonas aeruginosa) were determined by the RL/IS60 system, most (91.1%) of them were within the QC limits proposed by the National Committee for Clinical Laboratory Standards. The system was further assessed for a total of 162 clinical isolates, including E. coli, Citrobacter freundii, Enterobacter cloacae, Klebsiella pneumoniae, Serratia marcescens, Proteus mirabilis, Morganella morganii, P. aeruginosa, Haemophilus influenzae, S. aureus, coagulase-negative staphylococci, Enterococcus faecalis, Enterococcus faecium, and Streptococcus pneumoniae. Overall, there was 90.6% agreement between the RL/IS60 system and the reference microdilution method. Our results suggest that the RL/IS60 system provides rapid and reliable MICs of a variety of antimicrobial agents for clinical isolates as well as QC strains.
UR - http://www.scopus.com/inward/record.url?scp=0037232054&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0037232054&partnerID=8YFLogxK
U2 - 10.1128/JCM.41.1.279-284.2003
DO - 10.1128/JCM.41.1.279-284.2003
M3 - Article
C2 - 12517861
AN - SCOPUS:0037232054
VL - 41
SP - 279
EP - 284
JO - Journal of Clinical Microbiology
JF - Journal of Clinical Microbiology
SN - 0095-1137
IS - 1
ER -