Development of real-time x-ray microtomography system

H. Takano, M. Morikawa, S. Konishi, H. Azuma, S. Shimomura, Y. Tsusaka, S. Nakano, N. Kosaka, K. Yamamoto, Y. Kagoshima

Research output: Contribution to journalConference articlepeer-review

8 Citations (Scopus)

Abstract

We have developed a four-dimensional (4D) x-ray microcomputed tomography (CT) system that can obtain time-lapse CT volumes in real time. The system consists of a high-speed sample rotation system and a high-frame-rate x-ray imager, which are installed at a synchrotron radiation x-ray beamline. As a result of system optimization and introduction of a "zoom resolution" procedure, a real-time 4D CT movie with a frame rate of 30 was obtained with a voxel size of 2.5 μm using 10 keV x-rays.

Original languageEnglish
Article number012025
JournalJournal of Physics: Conference Series
Volume463
Issue number1
DOIs
Publication statusPublished - 2013
Externally publishedYes
Event11th International Conference on X-Ray Microscopy, XRM 2012 - Shanghai, China
Duration: 2012 Aug 52012 Aug 10

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Fingerprint Dive into the research topics of 'Development of real-time x-ray microtomography system'. Together they form a unique fingerprint.