Development of monoenergetic neutron calibration fields between 8 keV and 15 MeV

M. Baba, M. Takada, T. Iwasaki, S. Matsuyama, T. Nakamura, H. Ohguchi, T. Nakao, T. Sanami, N. Hirakawa

Research output: Contribution to journalArticlepeer-review

64 Citations (Scopus)

Abstract

For characterization and calibration of neutron dosimeters and spectrometers, monoenergetic neutron calibration fields have been developed at eight energy points between 8 keV and 15 MeV (8 and 27 keV, 0.25, 0.55, 1.0, 2.0, 5.0 and 15 MeV). Monoenergetic neutrons are obtained by the Sc(p, n) reactions for 8 and 27 keV, and by the 7Li(p, n), T(p, n), D(d, n) and T(d, n) reactions between 0.25 and 15 MeV. Care was taken to reduce background neutrons by parasitic reactions and the scattering of primary neutrons at the target. The neutron spectrum was characterized by the time-of-flight technique, and the neutron fluence was measured with two independent methods within ±5%. The field has been applied successfully for calibration and characterization of various neutron dosimeters and spectometers.

Original languageEnglish
Pages (from-to)115-123
Number of pages9
JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume376
Issue number1
DOIs
Publication statusPublished - 1996 Jun 21

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Instrumentation

Fingerprint

Dive into the research topics of 'Development of monoenergetic neutron calibration fields between 8 keV and 15 MeV'. Together they form a unique fingerprint.

Cite this