Development of apparatus for multiple energy X-ray holography at SPring-8

Kouichi Hayashi, Masao Miyake, Tomoaki Tobioka, Yasuhiro Awakura, Motohiro Suzuki, Shinjiro Hayakawa

Research output: Contribution to journalArticlepeer-review

24 Citations (Scopus)

Abstract

We developed an apparatus for multiple energy X-ray holography (MEXH) at third generation synchrotron facility, SPring-8. The combination system of a cylindrical lithium fluoride crystal and an avalanche photo diode in our apparatus enabled high countrate detection of the fluorescence from samples. From the measurements repeated 10 times, reproducible holographic undulations were obtained from the strontium titanate single crystal, although 10% intensity change of the incident beam occurred. This result revealed that the holographic data provided by our MEXH apparatus was highly stable to the variation of the incident beam intensity.

Original languageEnglish
Pages (from-to)1241-1244
Number of pages4
JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume467-468
Issue numberPART II
DOIs
Publication statusPublished - 2001

Keywords

  • Avalanche photodiode
  • Cylindrical lithium fluoride crystal
  • Multiple energy X-ray holography
  • Reproducibility
  • Strontium titanate
  • X-ray fluorescence holography

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Instrumentation

Fingerprint Dive into the research topics of 'Development of apparatus for multiple energy X-ray holography at SPring-8'. Together they form a unique fingerprint.

Cite this