Development of a high mass-resolution TOF-ERDA system for a wide mass range

W. Hong, S. Hayakawa, Kazushige Maeda, S. Fukuda, M. Yanokura, M. Aratani, K. Kimura, Y. Gohshi, I. Tanihata

Research output: Contribution to journalArticlepeer-review

14 Citations (Scopus)

Abstract

A high mass-resolution TOF-ERDA system installed at RILAC of RIKEN is presented. The target chamber was designed for TOF-ERDA. A charged-particle detector and a time detector for TOF measurements were installed in the chamber. The data-acquisition system, comprising CAMAC modules, is controlled by a personal computer. The analysis of the spectra is performed off-line. The mass resolution of the system for masses of around A = 28 was evaluated to be 1.9. H, C, N and O in a PIQ film could be measured simultaneously, and the peaks of C, N and O were separated.

Original languageEnglish
Pages (from-to)95-99
Number of pages5
JournalNuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
Volume124
Issue number1
DOIs
Publication statusPublished - 1997 Jan 1

Keywords

  • Light element analysis
  • TOF-ERDA

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Instrumentation

Fingerprint Dive into the research topics of 'Development of a high mass-resolution TOF-ERDA system for a wide mass range'. Together they form a unique fingerprint.

Cite this