Application of probe electrospray ionization-tandem mass spectrometry to ultra-rapid determination of glufosinate and glyphosate in human serum

Kiyotaka Usui, Eriko Minami, Yuji Fujita, Eito Kubota, Haruka Kobayashi, Tomoki Hanazawa, Tomohiro Yoshizawa, Yoshito Kamijo, Masato Funayama

Research output: Contribution to journalArticlepeer-review

4 Citations (Scopus)

Abstract

Glufosinate and glyphosate, which are non-selective herbicides that include an amino acid moiety in their structures, are frequently used worldwide to control unwanted vegetation. Unfortunately, these readily available herbicides are also used by people to commit suicide, and thus represent important chemicals of interest in the fields of clinical medicine and forensics. Because of the high water solubility of these herbicides, most analytical methods for their detection require a derivatization step, which results in longer analysis times. Therefore, derivatization-based methods do not currently contribute to judgements on treatment decisions in emergency medicine. In this study, we addressed this limiting factor by developing an ultra-rapid and simple analytical technique using a combination of probe electrospray ionization (PESI) and tandem mass spectrometry (MS/MS), which gives quantitative results within 0.3 min. Herbicide standards were added to human serum that was then subjected to analysis (N = 5 per concentration). The analysis was repeated daily over eight consecutive days. The limit of detection (LOD) was 0.59 μg/mL for glufosinate and 0.20 μg/mL for glyphosate. The limit of quantitation (LOQ), i.e., the lowest point on the calibration curves, was 1.56 μg/mL for both the herbicides. The matrix effects were observed at three different concentrations (between 95.7%–104% for glufosinate, and between 90.7%–95.7% for glyphosate). When applied to samples taken from actual poisoning cases (six samples for each herbicide), the present method gave almost the same quantitative values as those obtained by conventional high-performance liquid chromatography with fluorescence detection. Thus, we believe that PESI-MS/MS could emerge as a rapid diagnosis method in the clinical emergency field.

Original languageEnglish
Pages (from-to)175-181
Number of pages7
JournalJournal of Pharmaceutical and Biomedical Analysis
Volume174
DOIs
Publication statusPublished - 2019 Sep 10
Externally publishedYes

Keywords

  • Ambient mass spectrometry
  • Glufosinate
  • Glyphosate
  • Herbicide poisoning
  • Probe electrospray ionization
  • Rapid detection

ASJC Scopus subject areas

  • Analytical Chemistry
  • Pharmaceutical Science
  • Drug Discovery
  • Spectroscopy
  • Clinical Biochemistry

Fingerprint Dive into the research topics of 'Application of probe electrospray ionization-tandem mass spectrometry to ultra-rapid determination of glufosinate and glyphosate in human serum'. Together they form a unique fingerprint.

Cite this