Abstract
An ultra-precision instrument with concomitant micromanipulation techniques is designed and set up to measure the damage strength of a single biomimetic microcapsule. It can provide the capability of simultaneously measuring the applied force and resultant displacement of a single microcapsule, with maximum force range of 5mN, resolution of 0.1μN and ultimate traveling distance up to 12mm, resolution of 1nm, respectively. By armed high magnification side-view system, it can offer extra and withal valuable information for the supervened analyzing. The bursting force of urea-formaldehyde biomimetic microcapsules of diameter 65μm in glucose solution was measured by this technique. The microcapsule was burst when the deformation reached a value of 56.2% of its diameter and the corresponding resonant force is about 1700μN. The technique provides an effective means to characterize elastic properties of micro biomimetic capsules and compare mechanical strength of microcapsules made of different ingredients.
Original language | English |
---|---|
Pages (from-to) | 1730-1735 |
Number of pages | 6 |
Journal | Key Engineering Materials |
Volume | 297-300 III |
DOIs | |
Publication status | Published - 2005 Jan 1 |
Keywords
- Biomimetic microcapsules
- Bursting and deformation
- Micromanipulation
ASJC Scopus subject areas
- Materials Science(all)
- Mechanics of Materials
- Mechanical Engineering